June 2003
2-4
Troubleshooting and Maintenance
Procedure for Testing
the Recorder Output
Make sure the recorder(s) are connected as described in
Connecting
the Recorder Outputs
on page 1-4, and use the following steps to
perform a test:
1.
Use the arrow keys to move the pointer to the channel and
recorder you want to test, then press [YES].
2.
Enter a percentage between 0 and 125 to test.
Note:
If you do not enter a new value, the Series 1 defaults to the
previously entered value.
3.
Press [YES].The recorder pen should swing to the appropriate
value.
Note:
The recorder output depends on the recorder range (0-20 mA,
4-20 mA, 0-2 V).
4.
Repeat steps 1 through 3 for each channel and recorder you want
to test.
5.
To exit, press the DONE menu key until Main Menu appears on
the message line.