Installing Optional Features
1-21
June 2003
Performing an MH
Calibration Test/
Adjustment (cont.)
Calibration Procedure (cont.)
3.
Reprogram the meter with the new (corrected) low reference value
(if required) as described in
Entering High and Low Reference
Values
in Chapter 2 of the
Programming Manual
.
4.
Verify that the probe cable is not connected to the probe.
5.
Note the zero reference readings and verify that the readings are
now within ±0.0003 MH.
6.
Fill out a new high and low reference sticker with the final low
reference value and/or record the information on the Program
Information List provided in Appendix A of the
Startup Guide
.
Make sure you record the information below:
HIGH REF = ORIGINAL VALUE
LOW REF = NEW CORRECTED VALUE
ZERO REF = ORIGINAL RECORDED VALUE
7.
Reconnect the probe to the cable.
Note:
If cables are changed in any way, repeat this procedure for
maximum accuracy.
The Series 1 is now ready for operation. Proceed to Chapter 2,
Basic
Programming,
of the
Programming Manual
for instructions.