June 2003
1-8
Installing Optional Features
Connecting Pressure
Sensor Inputs
The Series 1 accepts either pressure transducers or pressure
transmitters with a 0/4-20 mA or 0 to 2-V output. Each type of sensor
is connected to the meter differently; therefore it is important to know
which type of pressure sensor you are using.
IMPORTANT:
The transducer must be supplied by GE Panametrics
or approved by GE Panametrics for use in this
circuit.
A pressure transducer is an electrically passive device that requires a
well-regulated excitation voltage or current. The transducer produces
a low level signal output (typically in the millivolt range) when
pressure is applied to it.
A pressure transmitter is an electrically active device containing
electronic circuits. A pressure transmitter requires some sort of power
source, such as a 24 VDC or 120 VAC. It produces an output signal
much larger than a pressure transducer’s in either current or voltage.
The more common pressure transmitters output current.
IMPORTANT:
The following connection information does not
pertain to the TF Series Probe.
To properly connect a pressure sensor, use the appropriate section
that follows.