
June 2003
1-12
Installing Optional Features
Connecting Pressure
Transmitters (cont.)
A. Connecting the Two-Wire or Loop-Powered Transmitter
Use a two-wire non-shielded cable to make connections to the
terminal block labeled AUX on the back of the electronics unit (refer
to Figure 1-9). Use Table 1-5 to make the proper pin connections to
the desired channel.
Note:
Twisted-pair cables work well with this circuit.
Once you complete the pressure connections, you must set switch
block S1 on the Series 1 channel card(s) for either current or voltage,
depending on the type of pressure sensor you are using (refer to
C.
Setting Input Switches
on page 1-14).
Figure 1-9: AUX Terminal Block
Table 1-5: Wire Connections for Two-Wire or Loop-
Powered Transmitters
Connect:
To AUX Terminal Block:
Positive Lead
pin +24V
Negative Lead
pin 2 (aux. input 2) or
pin 1 (aux. input 1)
AUX
Terminal Block
Содержание Moisture Image 1 Series
Страница 2: ...Moisture Image Series 1 Service Manual ...
Страница 8: ...Chapter 1 ...
Страница 31: ...Chapter 2 ...
Страница 61: ...Appendix A ...
Страница 94: ...June 2003 A 32 Application of the Hygrometer 900 901D1 Figure A 2 Moisture Content Nomograph for Liquids ...