Troubleshooting and Maintenance
2-7
June 2003
Procedure for Trimming
the Zero Value (cont.)
Figure 2-5: Recorder Test Menu
2.
Use the arrow keys to move the pointer to the recorder you want to
test and press [YES].
3.
Enter the % of scale. If your recorder is set up for 4 to 20-mA,
enter 0 and press [YES]. If your recorder is set up for 0 to 20-mA
or 0 to 2-V output, enter 5 and press [YES].
Note:
The recorders can not be trimmed to output a value of 0.00
mA/0.000 V due to the limits imposed by electronic noise. The
recorder output is typically 0.01 mA at zero output; therefore,
you should use 5% for the test value for 0 to 20-mA and 0 to 2-
V ranges.
RECORDER OUTPUT TEST
Ch Recorder A Recorder B
% of Scale
1 0 0
3 0 0
5 0 0
6 0 0
NOTE: Recorders SUSPENDED during test.
Recorder Test Menu
HELP TRIM --- DONE
2 0 0
4 0 0