Revision E
MAC
5500 resting ECG analysis system
B-3
2020299-020
Appendix B – Technical Specifications: Technical Specifications
Technical Specifications
Display
Item
Description
Type
264mm (10.4 in.) diagonal graphics backlit AM LCD
Resolution
640 x 480 pixels, with waveform enhancement
Displayed Data
Heart rate, patient name, ID, clock, waveforms, lead labels, speed, gain and filter settings,
warning messages, prompts and help messages
Computerized Electrocardiograph
Item
Description
Instrument Type
15 lead (14 channel) microprocessor-augmented, automatic electrocardiograph
Analysis Frequency
500 samples/s (sps)
ECG Storage
200 (maximum) on internal, non-volatile memory. Facility to archive ECG records on
removable media (SD card).
Digital Sampling Rate
4000 samples/s/channel
Analysis
Pediatric and vectorcardiography
Optional: 12SL analysis, HI-RES and PHI-RES late potential analysis
Pre-acquisition
Provides 10 s of instantaneous ECG acquisition
Dynamic Range
AC differential:
±5 mV
DC offset:
±320 mV
Resolution
4.88 µV/LSB @ 250 sps, 1.22 µV/LSB @ 500 sps
Frequency Response
–3 dB @ 0.01 to 150 Hz
Common Mode Rejection
>140 dB (123 dB with AC filter disabled)
Input Impedance
>10 M
@ 10Hz, defibrillator protected
Patient Leakage Current
<10 µA
Pace Detection
750 µV @ 50 µs duration, Orthogonal LA, LL and V6
Special Acquisition Functions
Disconnected lead detection, electrode impedance, AC noise, baseline wander, and
muscle tremor
Communication
MAC and MUSE system compatible RS-232
Optional: Modem, LAN, FAX, and wireless transmission, remote retrieval (remote query)
Содержание MAC 5500
Страница 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Страница 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Страница 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Страница 18: ...1 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction Service Information ...
Страница 19: ...Revision E MAC 5500 resting ECG analysis system 2 1 2020299 020 Equipment Overview 2 Equipment Overview ...
Страница 20: ...2 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview For your notes ...
Страница 54: ...2 36 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview Theory of Operation ...
Страница 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Страница 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Страница 73: ...Revision E MAC 5500 resting ECG analysis system 4 1 2020299 020 Troubleshooting 4 Troubleshooting ...
Страница 74: ...4 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Troubleshooting For your notes ...
Страница 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Страница 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Страница 136: ...5 32 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance Functional Checkout Procedures ...
Страница 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Страница 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Страница 146: ...6 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List Field Replaceable Units ...
Страница 147: ...Revision E MAC 5500 resting ECG analysis system A 1 2020299 020 Appendix A Abbreviations A Appendix A Abbreviations ...
Страница 148: ...A 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations For your notes ...
Страница 160: ...B 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications For your notes ...
Страница 168: ...C 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility For your notes ...
Страница 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Страница 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 179: ......