Revision E
MAC
5500 resting ECG analysis system
5-13
2020299-020
Maintenance: Disassembly Guidelines
Main CPU Board
Removal of CPU Board
NOTE
Before you begin, save the current System Setups to an SD card and print
System Setup report. This will be used to restore the system setups after
replacement of the CPU board.
1.
Remove the battery.
2.
Remove the top cover assembly following the procedures in
3.
Remove the display/keyboard assembly following the procedures in
4.
Disconnect all remaining cable connections to the main PCB. These include
cables to the following:
power supply
printhead
battery connect PCB
acquisition module cable
5.
Remove the COMM board.
6.
With a TORX driver, remove the mounting screws holding the main PCB in
place. They are located around the outside edges of the main PCB. Set screws
aside for mounting new board.
7.
Remove the harness cable.
8.
Lift the main PCB from the unit.
Reassembly of CPU Board
1.
Insert the new CPU board in place and mount using the screws set aside during
disassembly.
2.
Reassemble the top cover and display/keyboard assemblies by reversing the
steps for removal.
3.
Insert rear bezel into slot on back of MAC 5500 assembly as shown below.
Содержание MAC 5500
Страница 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Страница 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Страница 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Страница 18: ...1 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction Service Information ...
Страница 19: ...Revision E MAC 5500 resting ECG analysis system 2 1 2020299 020 Equipment Overview 2 Equipment Overview ...
Страница 20: ...2 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview For your notes ...
Страница 54: ...2 36 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview Theory of Operation ...
Страница 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Страница 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Страница 73: ...Revision E MAC 5500 resting ECG analysis system 4 1 2020299 020 Troubleshooting 4 Troubleshooting ...
Страница 74: ...4 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Troubleshooting For your notes ...
Страница 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Страница 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Страница 136: ...5 32 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance Functional Checkout Procedures ...
Страница 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Страница 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Страница 146: ...6 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List Field Replaceable Units ...
Страница 147: ...Revision E MAC 5500 resting ECG analysis system A 1 2020299 020 Appendix A Abbreviations A Appendix A Abbreviations ...
Страница 148: ...A 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations For your notes ...
Страница 160: ...B 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications For your notes ...
Страница 168: ...C 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility For your notes ...
Страница 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Страница 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 179: ......