Revision E
MAC
5500 resting ECG analysis system
4-9
2020299-020
Troubleshooting: Diagnostic Tests
5.
Select
Miscellaneous Setup
.
6.
Select the
System password
line and type the new password in the space.
7.
Press the
Enter
key.
8.
Select
Save Setup
from the
System Setup
menu.
9.
Select
To system
.
System Diagnostics Main Menu
Use the arrow pad control to highlight a menu item, then press the
Enter
key to
select it. The tests and test menus contain on-line prompts and/or instructions.
Display Tests
Speaker Test
Keyboard Test
Writer Tests
Battery Tests
Communication Tests
Acq. Module Tests
Analog I/O Tests
Floppy Drive Tests
Internal Memory Tests
SD Card Tests
Exit System Diagnostics
(reboots the system)
Display Tests
Run the screen
Display Tests
to verify that all the screen pixels are working and that
the brightness and contrast samples appear to be within normal range. There are no
screen display adjustments. The screen
Display Tests
are as follows.
Pixel Verification Test
Use the arrow pad control to move the bar across the screen and look for any
missing pixels on the display.
Press the
F1
key to turn on all of the pixels simultaneously.
Press the
Enter
key to exit the test.
Grey Scale Test Patterns
The first test pattern (used in manufacturing to verify the screen intensity) shows
two squares, one bright and one dim. Press any key to activate the next display.
The second test pattern shows the 32-color text palette (Various gray scale patterns
appear if the system does not have the color option enabled.). Check for problems
with the overall pattern. (If the system does not have the color option, various grey
scale patterns display.)
Press any key to exit the test.
Содержание MAC 5500
Страница 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Страница 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Страница 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Страница 18: ...1 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction Service Information ...
Страница 19: ...Revision E MAC 5500 resting ECG analysis system 2 1 2020299 020 Equipment Overview 2 Equipment Overview ...
Страница 20: ...2 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview For your notes ...
Страница 54: ...2 36 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview Theory of Operation ...
Страница 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Страница 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Страница 73: ...Revision E MAC 5500 resting ECG analysis system 4 1 2020299 020 Troubleshooting 4 Troubleshooting ...
Страница 74: ...4 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Troubleshooting For your notes ...
Страница 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Страница 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Страница 136: ...5 32 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance Functional Checkout Procedures ...
Страница 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Страница 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Страница 146: ...6 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List Field Replaceable Units ...
Страница 147: ...Revision E MAC 5500 resting ECG analysis system A 1 2020299 020 Appendix A Abbreviations A Appendix A Abbreviations ...
Страница 148: ...A 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations For your notes ...
Страница 160: ...B 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications For your notes ...
Страница 168: ...C 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility For your notes ...
Страница 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Страница 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 179: ......