C-4
MAC
5500 resting ECG analysis system
Revision E
2020299-020
Appendix C – Electromagnetic Compatibility: Electromagnetic Compatibility (EMC)
Guidance and Manufacturer's Declaration - Electromagnetic
Immunity
The MAC 5500 is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to ensure that the MAC 5500 is
used in such an environment.
NOTE
Ut is the AC mains voltage prior to application of the test level.
Immunity Test
EN 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
EN 61000-4-4
± 2 kV for power supply lines
±1 kV for input/output lines
± 2 kV for power supply lines
±1 kV for input/output lines
Mains power should be that of a typical
commercial or hospital environment.
Surge
EN 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN 61000-4-11
<5% Ut (>95% dip in U
t
)
for 0.5 cycles
40% Ut (60% dip in U
t
)
for 5 cycles
70% Ut (30% dip in U
t
)
for 25 cycles
<5% Ut (>95% dip in U
t
)
for 5 sec
<5% Ut (>95% dip in U
t
)
for 0.5 cycles
40% Ut (60% dip in U
t
)
for 5 cycles
70% Ut (30% dip in U
t
)
for 25 cycles
<5% Ut (>95% dip in U
t
)
for 5 sec
Mains power should be that of a typical
commercial or hospital environment. If the
user of the MAC 5500 requires continued
operation during power mains interruptions,
it is recommended that the MAC 5500 be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristics of a typical location
in a typical commercial or hospital
environment.
Содержание MAC 5500
Страница 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Страница 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Страница 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Страница 18: ...1 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction Service Information ...
Страница 19: ...Revision E MAC 5500 resting ECG analysis system 2 1 2020299 020 Equipment Overview 2 Equipment Overview ...
Страница 20: ...2 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview For your notes ...
Страница 54: ...2 36 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview Theory of Operation ...
Страница 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Страница 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Страница 73: ...Revision E MAC 5500 resting ECG analysis system 4 1 2020299 020 Troubleshooting 4 Troubleshooting ...
Страница 74: ...4 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Troubleshooting For your notes ...
Страница 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Страница 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Страница 136: ...5 32 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance Functional Checkout Procedures ...
Страница 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Страница 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Страница 146: ...6 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List Field Replaceable Units ...
Страница 147: ...Revision E MAC 5500 resting ECG analysis system A 1 2020299 020 Appendix A Abbreviations A Appendix A Abbreviations ...
Страница 148: ...A 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations For your notes ...
Страница 160: ...B 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications For your notes ...
Страница 168: ...C 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility For your notes ...
Страница 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Страница 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 179: ......