Revision E
MAC
5500 resting ECG analysis system
5-21
2020299-020
Maintenance: Disassembly Guidelines
5.
Enter the
IP Address, Subnet Mask,
and
Gateway
provided by the network
administrator
6.
Enter the
Port Number
used by the MUSE system.
7.
Select
Return
and press
Enter
.
The Network Settings window closes.
8.
On the Basic System menu, select Transmission and press Enter.
The
Transmission
window opens.
9.
Enter the following information.
10. Select
Return
and press
Enter
.
The
Transmission
window closes.
11. On the
Basic System
menu, select
Return
and press
Enter
.
The
Basic System
menu closes.
12. On the
System Setup
menu, select
Save Setup
>
To System
.
This saves the settings to the communication board. This should take
approximately 8 to 10 seconds. If it takes less than 8 seconds, the settings were
not saved successfully. Repeat a second time.
13. Cycle power on the MAC 3500/MAC 5500 system.
This ensures the network settings are reread from the communications board.
14. Wait at least 10 minutes and verify the MAC 3500/MAC 5500 can
communicate with the MUSE system.
Windows File Server 2000/2003 has an ARP cache that regenerates the MAC
Address/Physical Address table every 10 minutes. When you replace the
communications board, you change the MAC Address/Physical Address. By
waiting at least 10 minutes, you allow the MUSE operating system to
regenerate the MAC Address/Physical Address cache so the MAC 3500/MAC
5500 is able to communicate again with the MUSE system.
15. Perform all necessary functional checkouts.
Table 2. Transmission Information
Field
Value
Use IR for serial line
No
Serial line baud rate
115.2K
Default location
Ethernet (MUSE)
Содержание MAC 5500
Страница 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...
Страница 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...
Страница 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...
Страница 18: ...1 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction Service Information ...
Страница 19: ...Revision E MAC 5500 resting ECG analysis system 2 1 2020299 020 Equipment Overview 2 Equipment Overview ...
Страница 20: ...2 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview For your notes ...
Страница 54: ...2 36 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview Theory of Operation ...
Страница 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...
Страница 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...
Страница 73: ...Revision E MAC 5500 resting ECG analysis system 4 1 2020299 020 Troubleshooting 4 Troubleshooting ...
Страница 74: ...4 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Troubleshooting For your notes ...
Страница 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...
Страница 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...
Страница 136: ...5 32 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance Functional Checkout Procedures ...
Страница 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...
Страница 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...
Страница 146: ...6 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List Field Replaceable Units ...
Страница 147: ...Revision E MAC 5500 resting ECG analysis system A 1 2020299 020 Appendix A Abbreviations A Appendix A Abbreviations ...
Страница 148: ...A 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations For your notes ...
Страница 160: ...B 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications For your notes ...
Страница 168: ...C 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility For your notes ...
Страница 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...
Страница 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...
Страница 179: ......