Revision B
MAC
™
5000 resting ECG analysis system
4-13
2024917-010
Maintenance:
Disassembly Guidelines
Display/Keyboard Reassembly
1. Insert both flex cables through flex cable slots and position them as
shown.
2. Tilt the display/keyboard assembly to the left and with the roll pin of
the hinge (metal rod) parallel to the left hinge base, insert the rod
into the left hinge base and lower the display/keyboard assembly in
place.
3. Slide tabs into their mounting slots and set the display/keyboard
assembly in place.
4. Connect the three cables from the display/keyboard assembly to the
main PCB. Be sure to lift the locks up prior to attempting to insert
the cables into the connectors.
5. Slide the display hinge (metal rod) to the right until it locks into the
right hinge base.
6. Replace the hinge bracket with the two TORX screws removed
earlier.
7. Replace the screw and display ground at the left of the hinge rod.
8. Replace the two TORX mounting screws on the right side of
assembly.
Reassembly
1. Raise the display to the vertical position.
2. Make sure the bezel surrounding the rear panel connectors is in
place. Make sure the release mechanism for the Smartmedia card
functions properly.
3. Lower the top cover down around the display and set in position.
4. Snap the rear of the top cover in place and then, gently pulling on the
thin plastic strip at the front of the top cover, position it in place
under the keyboard assembly.
5. Replace the screws removed in disassembly.
45A
44A
Align Flex to Pin
Flex Cable
Slots
Display Ground
LCD Flex Cable
Backlight
Flex Cable
Backlight Flex Cable
LCD Flex Cable
Fasten to writer as shown
Roll Pin
Display Ground
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......