3-16
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Installation:
MAC 5000 ST Requirements and Configuration
MAC 5000 Configuration Requirements – At the
Main Menu
complete
the following in the order shown below:
Select
System Setup
,
Enter
System password
,
Exercise
Test
,
Inputs/Outputs
,
Change
Blood Pressure
to
Suntech
.
Ergoline - Model Ergoline 900
Connection Requirements – Use cable PN 2008110-001 to connect
from the MAC 5000 port
1
to the Ergoline serial port. Use cable
PN 2008115-001 to connect from the MAC5000
ANA/TTL
port to
the Ergoline QRS trigger input.
Device Configuration Requirements – See Ergoline 900 Operator’s
Manual.
MAC 5000 Configuration Requirements – At the
Main Menu
complete
the following in the order shown below:
Select
System Setup
,
Enter
System password
,
Exercise
Test
,
Inputs/Outputs
,
Change
Blood Pressure
to Ergoline Ergometer.
Compatible GE Medical Systems
Information
Technologies
Treadmills
Model T2000
Connection Requirements – Use cable PN 2007918-001 (T2000) to
connect from the MAC 5000 port
1
to the treadmill serial port.
Device Configuration Requirements – None.
MAC 5000 Configuration Requirements – Use the
Edit Protocol
application to set the protocol Test Type to
Treadmill in MPH
or
Treadmill in Km/H
for protocols that will be used with this treadmill.
Analog Treadmills
Connection Requirements – There are no cables available from GE
Medical Systems
Information
Technologies
to interface to analog
treadmills. The customer is responsible for making the appropriate
cable. Speed and grade signals for controlling analog treadmills are
available on pins 2 (Slow Analog Output) and 8 (Fast Analog Output)
of the
ANA/TTL
port. Pins 1, 4 and 5 are tied to ground.
Device Configuration Requirements – None.
MAC5000 Configuration Requirements – Use the Edit Protocol
application to set the protocol Test Type to
Analog Treadmill in MPH
or
Analog Treadmill in Km/H
for protocols that will be used with this
treadmill.
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......