5-16
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Troubleshooting:
Diagnostic Tests
Analog Input Test
Follow the instructions on screen to connect a DC power supply to the DC
input pins of the
ANA/TTL
connector. The voltage of the DC input
displays. If test fails, the main CPU board may need to be replaced.
DCOut Loopback Test
Follow the instructions on screen to connect the DC Outputs to the
Analog Inputs. The test sends all possible values out the DC Outputs
and confirms that the correct values are read from the Analog Inputs.
A pass/fail result displays. If either test fails, the main CPU board may
need to be replaced.
NOTE
Pin numbers refer to the
ANA/TTL
port.
ECGOut/QRSTrigger Loopback Test
Follow the instructions on screen to connect the ECG Output and QRS
Trigger Output to the Analog Inputs. The test sends all possible values
out the ECG Output and a square wave out the QRS Trigger Output. It
confirms that the correct values are read from the Analog Inputs. A
pass/fail result displays. If either test fails, the main CPU board may
need to be replaced.
NOTE
Pin numbers refer to the
ANA/TTL
port.
Floppy Drive Tests
Follow the instructions on screen. A read/write test is performed on a
formatted floppy disk and a pass/fail test result is displayed. Try another
disk if this test fails. If this test continues to fail, contact GE Medical
Systems-
Information
Technologies
for service.
NOTE
The following test and the resultant values are for manufacturing
use only and NOT intended for service of this device.
A head radial alignment and Azimuth alignment test is performed using
an Accurite test disk (part number is displayed on screen). Alignment
test values will be displayed.
Diskette Format Failure
Unformatted diskettes may cause the message “
Please insert a data
diskette”
to appear and not allow the MAC 5000 to force a format.
Remove the diskette from the MAC 5000 and format the diskette on a
PC. After formatting, try it on the MAC 5000 again. If it fails again,
replace the diskette and repeat the procedure.
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......