4-24
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Maintenance:
Domestic Electrical Safety Tests
Domestic Electrical Safety Tests
AC Line Voltage Test
This test verifies that the domestic wall outlet supplying power to the
equipment is properly wired. For international wiring tests, refer to the
internal standards agencies of that particular country.
120 VAC, 50/60 Hz
Use a digital voltmeter to check the voltages of the 120-volt AC wall
outlet (dedicated circuit recommended). If the measurements are
significantly out of range, have a qualified electrician repair the outlet.
The voltage measurements should be as follows:
1. 120 VAC (
±
10 VAC) between the line contact and neutral and
between the line contact and ground.
2. Less than 3 VAC between neutral and ground.
240 VAC, 50/60 Hz
Use a digital voltmeter, set to measure at least 300 VAC, to check the
voltages of the NEMA 6-20R, AC wall outlet (dedicated circuit
recommended). If the measurements are significantly out of range, have
a qualified electrician repair the outlet. The voltage measurements
should be as follows:
1. 120 VAC (
±
10 VAC) between either “hot” contact and ground.
2. 210 to 230 VAC between the two “hot” contacts.
47A
48A
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......