Revision B
MAC
™
5000 resting ECG analysis system
3-17
2024917-010
Installation:
MAC 5000 ST Requirements and Configuration
Configure pin 2 on the
ANA/TTL
port by selecting the following:
System Setup
,
Exercise Test
,
Inputs/Outputs
, and
set
Slow Analog Output
to
Workload
.
Configure pin 8 on the
ANA/TTL
port by selecting the following:
System Setup
,
Exercise Test
,
Inputs/Outputs
, and
Set
Fast Analog Output
to
Workload
.
Bicycle Ergometers
Ergoline 800/900, Lode Ergometer
Connection Requirements – Use cable PN 2008109-001 (Ergoline 800),
PN 2008114-001 (Ergoline 900), or PN 2007981-001 (Lode Ergometer), to
connect from the MAC 5000
ANA/TTL
port to the ergometer analog
control port.
NOTE
For any other ergometer, the customer is responsible for making the
appropriate cable.
Device Configuration Requirements – Refer to ergometer Operator's
Manual.
MAC 5000 Configuration Requirements – Use the
Edit Protocol
application to set the protocol Test Type to
Ergometer in Watts
or
Ergometer in KPM
for protocols that will be used with this ergometer.
Configure pin 2 on the
ANA/TTL
port by selecting the following:
System Setup
,
Exercise Test
,
Inputs/Outputs
, and
Slow
Analog Output to
Workload
, or
Configure pin 8 by selecting:
System Setup
→
Exercise Test
→
Inputs/Outputs
→
Fast Analog Output to Workload
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......