2-6
MAC
™
5000 resting ECG analysis system
Revision B
2024917-010
Equipment Overview:
Theory of Operation
Theory of Operation
Overview / Block Diagram
The MAC 5000 CPU board contains all of the circuitry for the MAC 5000
electrocardiograph except for the line power supply, acquisition module,
keyboard and display. Although MAC 5000 runs software derived from
products based on the Max-1 architecture (running on the C-Exec
operating system), it has almost nothing in common with that hardware
family.
In a nutshell, the board contains the following:
64 MB SDRAM (holds both code and data)
32 MB SDRAM acts as video frame memory
32 MB NAND Flash (holds FPGA configuration and system code)
32 MB NAND Flash for ECG Record storage
128 KB Boot Data Flash (holds primary boot image)
CRT video DACs
External 12 Volt Power Switch
Acquisition Module Transceiver / Power Switch
Printhead Power Switches and Pixel Test Circuit
Daughter Board Interface which support serial ports, USB and
PC card.
Switch Mode Power Supplies
3.3 Volt for Logic, LCD
5 Volt for Logic, Printer, Floppy Drive
12 Volt for LCD backlight, External Com Port Power
Battery Charger
-12 Charge Pump for Analog Circuits
Linear Power Supplies
1.8 Volt (AT91RM9200 Core and FPGA Core)
2.5 Volt Reference
3.3 Volt for System Supervisor (Moe Stooge)
12 Volt for Analog Circuits
Crystals / Clocks
24 MHz Oscillator for FPGA
32.768 Khz Real Time Clock for Super IO chip.
32.768 Khz (AT91RM9200)
18.432 Mhz (AT91RM9200)
4 Mhz (3 devices, 1 for each Stooge)
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......