Revision B
MAC
™
5000 resting ECG analysis system
C-3
2024917-010
Appendix C – Electromagnetic Compatibility:
Electromagnetic Compatibility (EMC)
Electromagnetic Compatibility (EMC)
Changes or modification to this system not expressly approved by
GE Medical Systems could cause EMC issues with this or other
equipment. This system is designed and tested to comply with applicable
regulation regarding EMC and needs to be installed and put into service
according to the EMC information stated as follows.
WARNING
Use of portable phones or other radio frequency (RF)
emitting equipment near the system may cause
unexpected or adverse operation.
WARNING
The equipment or system should not be used adjacent to,
or stacked with, other equipment. If adjacent or stacked
use is necessary, the equipment or system should be
tested to verify normal operation in the configuration in
which it is being used.
Guidance and Manufacturer's Declaration - Electromagnetic
Emissions
The MAC 5000 is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure
that the MAC 5000 is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR11
Group 1
The equipment uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions
CISPR11
Class A
The equipment is suitable for use in all
establishments including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic Emissions
EN 61000-3-2
Class A
Voltage fluctuations/
Flicker emissions
EN 61000-3-3
Complies
Содержание MAC 5000
Страница 9: ...Revision B MAC 5000 resting ECG analysis system 1 1 2024917 010 1 Introduction ...
Страница 10: ...1 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 21: ...Revision B MAC 5000 resting ECG analysis system 2 1 2024917 010 2 Equipment Overview ...
Страница 22: ...2 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 59: ...Revision B MAC 5000 resting ECG analysis system 3 1 2024917 010 3 Installation ...
Страница 60: ...3 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 77: ...Revision B MAC 5000 resting ECG analysis system 4 1 2024917 010 4 Maintenance ...
Страница 78: ...4 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 103: ...Revision B MAC 5000 resting ECG analysis system 5 1 2024917 010 5 Troubleshooting ...
Страница 104: ...5 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 135: ...Revision B MAC 5000 resting ECG analysis system 6 1 2024917 010 6 Parts List ...
Страница 136: ...6 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 143: ...Revision B MAC 5000 resting ECG analysis system A 1 2024917 010 A Appendix A Abbreviations ...
Страница 144: ...A 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 155: ...Revision B MAC 5000 resting ECG analysis system B 1 2024917 010 B Appendix B Technical Specifications ...
Страница 156: ...B 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 163: ...Revision B MAC 5000 resting ECG analysis system C 1 2024917 010 C Appendix C Electromagnetic Compatibility ...
Страница 164: ...C 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 For your notes ...
Страница 171: ...Revision B MAC 5000 resting ECG analysis system Index 1 2024917 010 Index ...
Страница 172: ...Index 2 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 174: ...Index Index 4 MAC 5000 resting ECG analysis system Revision B 2024917 010 ...
Страница 175: ......