Chapter 7
Field Replaceable Unit (FRU) Replacement
Procedures
7.1
Replacing the Parts of the Lullaby
TM
Warmer
WARNING:-
Before replacing the Field Replacing Unit (FRU)s always do the following:
1.
Switch off the power.
2.
Set the castor brakes.
3.
Remove the cylinders if they are mounted on to the Lullaby
TM
Warmer.
4.
Always use anti-static material and wear anti-static wrist strap, safety glasses,
safety shoes and safety gloves while dismantling the Lullaby
TM
Warmer to avoid
possible damage due to static electricity and sharp edges.
7.1.1
Replacing the Hood Exterior (FRU Kit No: 2043936-001)
WARNING: -
The heater assembly and the protective grid may still be hot if you dismantle the unit
immediately after it has been in use.
CAUTION: -
Do not touch the inside of the heater reflector. Imprints of fingers may reduce the
output of the heater’s reflective surface. The temperature of the heater can be
determined by holding the skin temperature probe against the heater.
Note
: Always perform the system checkout procedure and electrical safety check before returning
the Lullaby
TM
Warmer to use after performing repair.
Illustration
Figure 81:
The heater hood with the components separated
Содержание Lullaby Warmer
Страница 12: ...This page is left blank intentionally ...
Страница 138: ...112 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer Hood Assembly Continued ...
Страница 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Страница 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Страница 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...