Lullaby
TM
Warmer 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 15
Over voltage and undervoltage indications.
Illustration
Description
Figure 21:
OVR
Figure 22:
UVR
If the applied mains voltage exceeds the
rated voltage by 10% for atleast 1
minute, “OVR” is displayed on the set
temperature display as seen in Figure 21.
The heating shall be cut off when “OVR” is
displayed on the display
WARNING
: Do not leave baby unattended
while “ovr” is displayed as heater will be
turned off.
If the applied voltage is lower than the
rated voltage by 10% for atleast 1
minute, “UVR” is displayed on the set
temperature display as seen in Figure 22.
In this condition, temperature monitoring
and all other warmer functions are
present.
Figure 23:
Heat OFF Indicator
Heat OFF indication:
When the hood is rotated to either of the sides
the heat level bar graph turns off and Heat
OFF indicator flashes.
NOTE: In the Heat OFF condition, audible
alarms are absent.
In this condition, temperature monitoring and
all other warmer functions are present.
Heat OFF
Indicator
Содержание Lullaby Warmer
Страница 12: ...This page is left blank intentionally ...
Страница 138: ...112 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer Hood Assembly Continued ...
Страница 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Страница 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Страница 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...