Lullaby
TM
Warmer 2042376-001 Rev G ©2012 General Electric Company. All rights reserved. 35
Symptom
Likely causes
Action
"CFg" is displayed on
Control
Temperature
display.
The configuration switch is
configured wrongly.
Ensure the configuration switch SW1
on the PWA is configured as per the
table.
B
A
230V
OFF
OFF
115V
ON
OFF
100V
OFF
ON
NOTE: For Warmer with SMPS and
LED observation lamp
Heater is cut off when
Hood is at center position
with system failure LED
blinking without alarm.
Swivel sensing cable not
connected.
Connect Swivel sensing cable
Swivel sensing cable is
defective.
Replace Swivel sensing cable
2044356-001 in FRU 2043950-001
NOTE: For Warmer with halogen
lamp and transformer.
Replace Swivel sensing cable
2060526-001 in FRU 2062276-001
NOTE: For Warmer with SMPS and
LED observation lamp.
Heater is cut off and the
bargraph is not turned on
when
the
hood
is
swiveled back to zero
position.
Limit
switch
is
not
actuated.
Gently adjust the hood near zero
position.
Содержание Lullaby Warmer
Страница 12: ...This page is left blank intentionally ...
Страница 138: ...112 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer Hood Assembly Continued ...
Страница 139: ...LullabyTM Warmer 2042376 001 Rev G 2012 General Electric Company All rights reserved 113 ...
Страница 140: ...114 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...
Страница 150: ...124 2042376 001 Rev G 2012 General Electric Company All rights reserved LullabyTM Warmer ...