8-30
Section 8-3 - Keyboard Rear Cover Assy (FRU P/N: 5315122)
8-3-1
Probe Holder (FRU P/N: 5168500) / Gel Bottle Holder (FRU P/N: 5168636)
This is a description on how to remove and replace the Probe holder Assembly.
8-3-1-1
Needed Manpower
•
1 person, 15 minutes
8-3-1-2
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-3-1-3
Removal procedure
1) Pull the probe holder out by slightly pushing it upwards with your hands.
Refer
Figure 8-32 Probe Holder
8-3-1-4
Mounting procedure
1.) Install the new parts in the reverse order of removal.
8-3-1-5
Functional Checkout Procedure
Functional Test
Debrief Script
Do Visual Inspection and ensure
no scratches observed and
fitment is done properly
Service Manual Direction 5308917-100, Section 8-3-2. Equipment passes all required
tests and is ready for use.
Gel Holder
Probe Holder