GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
4-24
Section 4-3 - General Procedure
4-3-10
Basic Measurements
NOTE:
The following instructions assume that you first scan the patient and then press
Freeze
.
4-3-10-1
Distance and Tissue Depth Measurements
1.) Press
MEASURE
once; an active caliper displays.
2.) To position the active caliper at the start point (distance) or the most anterior point (tissue
depth), move the
TRACKBALL
.
3.) To fix the start point, press
Set
. The system fixes the first caliper and displays a second active
caliper.
4.) To position the second active caliper at the end point (distance) or the most posterior point
(tissue depth), move the
TRACKBALL
.
5.) To complete the measurement, press
SET
. The system displays the distance or tissue depth
value in the measurement results window.
Before you complete a measurement:
To toggle between active calipers, press
MEASURE
.
To erase the second caliper and the current data measured and start the measurement again,
press
CLEAR
once.
NOTE:
To rotate through and activate previously fixed calipers, adjust
CURSOR SELECT
.
NOTE:
After you complete the measurement, to erase all data that has been measured to this point,
but not data entered onto worksheets, press
CLEAR
.
4-3-10-2
Circumference/Area (Ellipse) Measurement
1.) Press
MEASURE
once; an active caliper displays.
2.) To position the active caliper, move the
TRACKBALL
.
3.) To fix the start point, press
SET
. The system fixes the first caliper and displays a second active
caliper.
4.) To position the second caliper, move the
TRACKBALL
.
5.) Adjust the
ELLIPSE
; an ellipse with an initial circle shape appears.
6.) To position the ellipse and to size the measured axes (move the calipers), move the
TRACKBALL
.
7.) To increase the size, rotate the
ELLIPSE
button clockwise. To decrease the size, contrarotate
the
ELLIPSE
button.
8.) To toggle between active calipers, press
MEASURE
.
9.) To complete the measurement, press
SET
. The system displays the circumference and area
in the measurement results window.
Before you complete a measurement:
-
To erase the ellipse and the current data measured, press
CLEAR
once. The original caliper
is displayed to restart the measurement.
-
To exit the measurement function without completing the measurement, press
CLEAR
a
second time.
4-3-10-3
Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.
Содержание LOGIQ P3 PRO
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software...