GE PROPRIETARY
TO
GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
Chapter 1 Introduction
1-7
“Equipotentiality” indicates the terminal
to be used for connecting equipotential
conductors when interconnecting
(grounding) with other equipment.
Connection of additional protective
earth conductors or potential
equalization conductors is not
necessary in most cases and is only
recommended for situations involving
multiple equipment in a high-risk patient
environment to provide assurance that
all equipment is at the same potential
and operates within acceptable leakage
current limits. An example of a high-
risk patient would be a special
procedure where the patient has an
accessible conductive path to the heart
such as exposed cardiac pacing leads.
Rear of console
Alternating Current symbol is in
accordance with IEC 60878-01-14.
Rear Panel, Rating Plate, Circuit
breaker label of console and front panel
(if applicable).
Date of manufacture.
The date could be a year, year and
month, or year, month and day, as
appropriate. See ISO 8601 for date
formates.
Rating Plate
Catalog or model number.
Rating Plate
Serial number
Rating Plate
Direct Current.
For products to be powered from a DC
supply.
Rating Plate
Type CF Defib-Proof Applied Part (heart
in the box with paddle) symbolis in
accordance with IEC 60878-02-06.
ECG Module
Table 1-6 Warnings
LABEL/SYMBOL
PURPOSE/MEANING
LOCATION
Содержание LOGIQ P3 PRO
Страница 2: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 1 2...
Страница 9: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL vii...
Страница 98: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 3 32 Section 3 10 Paperwork...
Страница 194: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 7 26 Section 7 6 Troubleshooting...
Страница 294: ...GE PROPRIETARY TO GE DIRECTION 5308917 100 REVISION 8 LOGIQ P3 SERVICE MANUAL 8 100 Section 8 14 Loading Software...