GE
PROPRIETARY TO GE
D
IRECTION
5308917-100, R
EVISION
8
LOGIQ P3 S
ERVICE
M
ANUAL
Chapter 8 Replacement Procedures
8-91
8-13-1
CWD Assy (FRU No: 5124847)
Purpose: This is a description on how to remove and replace the CWD Assy.
8-13-1-1
Tools
•
Common phillips screwdrivers
•
Allen/Unbraco wrench
8-13-1-2
Needed Manpower
•
1 person, 30 m travel
8-13-1-3
Preparations
•
Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
Figure 8-92 Removal Procedure of cwd Refer to
1.) Unscrew two screws(1-2).
2.) Disconnect the CWD assy.
8-13-1-4
Removal procedure
.
1.) Unscrew two screws (1-2) to remove the CWD cover. see
]
2.) Connect the CWD assy to RX64. [See
]
3.) Press the CWD assy tightenedly, refer to
4.) Screw two screws (1-2) to fix the CWD assy on RX64 assy., see