Optimizing Spectral Doppler
LOGIQ 7 Online Help
5
-
101
Direction 2392536-100 Rev. 1
Wall Filter
Description
Insulates the Doppler signal from excessive noise caused from
vessel movement.
Adjusting
To increase/decrease, select
Wall Filter
, then adjust
Wall Filter
on the Touch Panel (PRF and Wall Filter share the same
control. Press the control to toggle between PRF and Wall
Filter). Each adjustment cycles you to the next setting.
Values
Values vary, depending upon the probe and application. The
current value displays on the Touch Panel and the monitor. Wall
Filter values are returned to the factory or user preset value
when you change the following: Probe, Exam Category, Exam
Calcs, New Patient.
Benefits
Gets rid of excess, unnecessary information. Cleans out low
level noise above and below the baseline so you don't see or
hear it on the spectrum.
Affect on other
controls
Wall filter may be changed by changes to the velocity scale.
Baseline
Description
Adjusts the baseline to accommodate faster or slower blood
flows to eliminate aliasing.
Adjusting
Baseline adjusts the point in the spectrum where the velocity
trace is at zero. The default baseline is at the midpoint of the
spectrum. To shift the baseline, adjust
Baseline
on the Touch
Panel.
The baseline displays as a solid line running across the
spectrum. The baseline is raised and lowered in equal
increments, depending on the current Doppler scale factor. The
control does not wrap when the maximum baseline shift (in
either direction) has been reached.
Values
12% increments, with 0 being the center of the display and
+100% being the top edge of the display and -100% being the
bottom edge of the display. Baseline values vary by probe and
application and are returned to the factory or user preset value
when you change: Probe, Exam Category, Exam Calcs, or New
Patient.
Benefits
Unwraps the alias. Rearranges the velocity scale without
changing the velocity scale. Readjusts the positive and negative
velocities limit without changing the total velocity range.
Содержание LOGIQ 7
Страница 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Страница 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 400: ...General Measurements and Calculations 7 94 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 699: ...LOGIQ 7 Basic User Manual 14 1 Direction 2392206 100 Rev 1 Chapter 14 ReportWriter Describes how to generate reports ...
Страница 741: ...LOGIQ 7 Basic User Manual 15 1 Direction 2392206 100 Rev 1 Chapter 15 Recording Images Describes how to record images ...
Страница 779: ...Presets LOGIQ 7 Basic User Manual 16 3 Direction 2392206 100 Rev 1 Overview continued Figure 16 1 Utility Touch Panel ...
Страница 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 983: ...GE Medical Systems ...
Страница 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...