Doppler Display
Time zero (the start of the trace) appears on the left side of the
graph. As time progresses, the trace moves to the right. The
baseline of the graph (representing zero velocity, zero frequency
shift, or no detected flow), appears as a solid line running
horizontally across the display. By convention, movement
toward the probe is positive and movement away from the probe
is negative. Positive frequencies or velocities appear above the
baseline. Negative frequencies or velocities appear below the
baseline.
Typically, blood flow is not uniform but is composed of a mix of
blood cells moving at different velocities and in different
directions. Thus, the display is composed of a spectrum as gray
scale values. Strong signals are displayed as bright while weak
signals are displayed as varying shades of gray.
HPRF is invoked when you are operating in PW Doppler Mode
and conditions activate HPRF (when the velocity scale factor or
sample volume gate depth exceeds certain limits). When HPRF
is active, multiple sample volume gates appear along the
Doppler mode cursor. Doppler information can be received from
any of the multiple sample volume gates. The Doppler signals
from all the gates are added together and displayed in one
spectrum.
Information about the PW Doppler display is automatically
written on the screen and updated when scanning parameters
are changed.
This chapter includes:
•
A discussion of PW Doppler.
•
Activating Pulsed Wave Doppler.
•
Optimizing the Doppler spectrum.
Imaged Doppler
When you view the PW Doppler spectrum and a B-Mode image
with a 3.5 MHz probe, this means that you are viewing the
image at 3.5 MHz and `Dopplering' at 2.5 MHz.
Содержание LOGIQ 7
Страница 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Страница 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 400: ...General Measurements and Calculations 7 94 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 699: ...LOGIQ 7 Basic User Manual 14 1 Direction 2392206 100 Rev 1 Chapter 14 ReportWriter Describes how to generate reports ...
Страница 741: ...LOGIQ 7 Basic User Manual 15 1 Direction 2392206 100 Rev 1 Chapter 15 Recording Images Describes how to record images ...
Страница 779: ...Presets LOGIQ 7 Basic User Manual 16 3 Direction 2392206 100 Rev 1 Overview continued Figure 16 1 Utility Touch Panel ...
Страница 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 983: ...GE Medical Systems ...
Страница 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...