Optimizing the Image
5
-
104
LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Dynamic Range
Description
Dynamic range controls how echo intensities are converted to
shades of gray, thereby increasing the range of contrast you can
adjust.
Adjusting
You access Doppler dynamic range from the Doppler Mode
Touch Panel. Each click cycles you to the next setting. To
increase/decrease, press
Dynamic Range
.
Values
The settings cycle in 4 dB increments. The current value
displays on the Touch Panel. Dynamic Range values vary by
probe and application and are returned to the factory or user
preset value when you change the following: Probe, Exam
Category, Exam Calcs, or New Patient.
Benefits
Optimizes the image's texture and smoothness by increasing
the amount of gray scale. You decrease the dynamic range to
clean out the spectrum to determine if the broadening is actual
or not. You increase the dynamic range to fill in the spectrum,
according to your personal preference.
Affect on other
controls
Dynamic range operates only in realtime, not in freeze, CINE,
CINE Timeline, or VCR playback and affects Doppler gain.
Spectral Trace (Trace Method)
Description
Traces the average mean and peak velocities in realtime images
and spectral broadening in frozen images.
Adjusting
Select the spectral trace method from
Trace Method
.
To activate the Spectral Trace Touch Panel, press
To get a peak trace, click MAX TRACE. A green trace displays
on the spectrum.
To get a mean trace, click MEAN TRACE. A blue trace displays
on the spectrum.
Selecting Trace
Position
SELECT TRACE lets you indicate whether cardiac cycle
detection, and therefore cardiac cycle identification, should use
peak timeline data above, below, or above and below
(composite) the baseline.
Benefits
Broadening may indicate turbulence.
Содержание LOGIQ 7
Страница 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Страница 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 400: ...General Measurements and Calculations 7 94 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 699: ...LOGIQ 7 Basic User Manual 14 1 Direction 2392206 100 Rev 1 Chapter 14 ReportWriter Describes how to generate reports ...
Страница 741: ...LOGIQ 7 Basic User Manual 15 1 Direction 2392206 100 Rev 1 Chapter 15 Recording Images Describes how to record images ...
Страница 779: ...Presets LOGIQ 7 Basic User Manual 16 3 Direction 2392206 100 Rev 1 Overview continued Figure 16 1 Utility Touch Panel ...
Страница 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 983: ...GE Medical Systems ...
Страница 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...