Optimizing B-Mode
LOGIQ 7 Online Help
5
-
7
Direction 2392536-100 Rev. 1
Gain
Description
B-Mode Gain increases or decreases the amount of echo
information displayed in an image. It may have the effect of
brightening or darkening the image if sufficient echo information
is generated.
Adjusting
Gain values vary depending on the probe; they are not
associated with a particular position of the knob.
NOTE:
TruAccess allows you to change the gain on a frozen image.
B-Mode gain is independent of M-Mode and Doppler and Color
Flow Gain. Changing the gain while in M-Mode does not affect
the B-Mode image gain.
To decrease/increase, rotate
Gain
.
Values
Gain displays on the monitor in Gn (dB). Gain increments by 2
dB. Maximum gain varies by probe. The current gain value is
displayed. Gain values vary by probe, application, and
frequency setting.
NOTE:
Maximum gain is factory preset to an optimum setting to
eliminate noise in the display.
Gain values are returned to the factory or user preset value
when you change the following: Probe, Exam Category/Exam
Calc, New Patient, or Frequency.
Benefits
Gain allows you to balance echo contrast so that cystic
structures appear echo-free and reflecting tissue fills in.
Affect on other
controls
After you adjust the Power Output, you may need to adjust the
gain. Generally speaking, if you increase the Power Output, you
need to decrease the gain; if you decrease the Power Output,
you need to increase the gain. Gain and TGC interact by adding
together.
Bioeffects
Gain has no affect on Power Output. However, with increased
gain, the power output level can usually be reduced to produce
an equivalent image quality.
NOTE:
Always optimize gain before increasing the Power Output.
Содержание LOGIQ 7
Страница 4: ...i 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 This page intentionally left blank ...
Страница 16: ...i 14 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 28: ...Introduction 1 12 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 140: ...Preparing for an Exam 4 28 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 306: ...Scanning Display Functions 6 42 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 400: ...General Measurements and Calculations 7 94 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 418: ...Abdomen and Small Parts 8 18 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 644: ...Cardiology 10 128 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 688: ...Urology 12 8 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 698: ...Pediatrics 13 10 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 699: ...LOGIQ 7 Basic User Manual 14 1 Direction 2392206 100 Rev 1 Chapter 14 ReportWriter Describes how to generate reports ...
Страница 741: ...LOGIQ 7 Basic User Manual 15 1 Direction 2392206 100 Rev 1 Chapter 15 Recording Images Describes how to record images ...
Страница 779: ...Presets LOGIQ 7 Basic User Manual 16 3 Direction 2392206 100 Rev 1 Overview continued Figure 16 1 Utility Touch Panel ...
Страница 882: ...Customizing Your System 16 106 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...
Страница 928: ...Probes and Biopsy 17 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 974: ...User Maintenance 18 46 LOGIQ 7 Online Help Direction 2392536 100 Rev 1 ...
Страница 983: ...GE Medical Systems ...
Страница 984: ...1 2 LOGIQ 7 Basic User Manual Direction 2392206 100 Rev 1 ...