Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5314528-2-1EN, Revision 1
5.3 Run Interconnection cables through the Conduit
Use the IVUS PIM (#806365003) as reference to the procedures below.
•
Unpack , and check the completeness of the Preinstallation kit :
•
1pcs - Custom PIM cable 30 m (98 ft, 5 in) - Volcano ID :806452001
•
2pcs - CAT5 cable 30 m (98 ft, 5 in) - Volcano ID :803391001
•
1pcs - Custom Twin axial cable - Volcano ID :806380001
•
1pcs – Ground Cable, 30 m (98 ft, 5 in) s5i – Volcano ID :806889001
•
1pcs - IVUS preinstallation Manual - Volcano ID :806365003
•
Install 806889001 Ground Cable through the Control Booth conduit (see note)
•
Install the cables through the IVUS conduit from the Control Room end :
•
806452001 Room - PIM cable, round shaped connector end to the Procedures
•
803391001 - Ethernet cable (1pcs)
•
Install the optional cables through the IVUS conduit from the Control Room end :
•
803391001 - Ethernet cable if 2 IVUS remote control items will be installed 1 IVUS
Control Station and 1 IVUS Joystick)
•
806380001 - ECG cable in case, if NON Mac-Lab ECG will be installed
NOTE:
For the ground cable, see the IVUS service manual for special instructions concerning
cable running and cutting .
NOTE:
If the installation has the MaC-Lab ECG in the configuration, the ECG cable shall not be run
through the conduit, it will be used in the Control Room to connect Mac-Lab and IVUS PCs.
NOTE:
You should consider running the ECG cable through the conduit in case, the customer
requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future expansion
options.
NOTE:
Same applies to the 2nd Ethernet cable, if a later extension of the IVUS remote controls
expected.
•
Ensure sufficient cable length on the Procedures Room end of the cables
according to the table below
Cable
Free length from the conduit exit
PIM cable
2.4 m (8 ’)
Control Station cable
2.1 m (7 ’)
Joystick cable
2.7 m (9 ’ )
ECG cable
Depends on the ECG location
NOTE:
The incorrect cable length might cause reliability issues during the operation !
Make sure, that the out of conduit parts of the cables with the connectors are properly
protected during the rest of the installation
170