Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5314528-2-1EN, Revision 1
2.3.3.2 2285646 – Through bolts without insert
Table 5-9:
Item
Name
Part #
Description
Quan.
Notes
Bolt, Hex
2296892
Through Bolt; M20 - 500-400
12 bolts for LC Positioner and 4 bolts
for Table
16 pc
Through Bolt mounting
method hardware
Washer, Flat
99142204
Special Flat Washer for Through
Bolts; one for each bolt
16 pc
same as above
Plate
2290941
Special Steel Spacer Plate; 4 in. x
4 in. (102 mm x 102 mm); one for
each bolt
16 pc
same as above
Nut, Hex
99141607
Hex Nut M20 STL galvanized two
for each bolt
32 pc
same as above
Grout assy
2285055
- 10 kg Powdered Mortar Ardex K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
1 kit
Used in constructing LC
Positioner grout dam
Dowel
2290937
Wood Dowel; 24 mm diameter
12 pc
Chapter 5 Physical Characteristics
141
Содержание Innova 2100-IQ
Страница 219: ......