Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5314528-2-1EN, Revision 1
3 Grounding
A vascular lab is a critical care area and requires a special grounding system for patient safety.
An equipotential grounding system is recommended for meeting patient safety requirements.
Reference: For general system grounding requirements and information on establishing an
equipotential grounding system, refer to: Grounding Continuity – Job card in the Installation
manual.
All shielded cables shall have a connection to ground at each extremity without
regard to the grounding to the scheme below.
Metallic covers on cable connectors shall be tested to ground.
Chapter 6 Electrical Connections
163