Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-1-3
Mounting Procedure
1.) Install the LCD monitor cover parts in the reverse order of removal procedure. Refer to the
"Monitor cover set" on page 8-5
.
2.) Put the label on the rear of the LCD monitor.
Table 8-2
Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394141, Rev 1+, Section 8-2-1. Equipment
passed all required tests and is ready for use. “
LCD plastic filter Function Validation
Figure 8-3 Putting the label on the rear of the LCD monitor