Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-111
8-2-49
Transbox Assy 110V/220V
8-2-49-1
Tools
•
Common pilIips screwdrivers
•
Allen/Unbraco wrench
8-2-49-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-49-3
Removal procedure
•
To detached the AC Power assy, All open 4 pcs console cover that are left, right, front, rear cover
1.) Remove side left cover and right side cover. Refer to the
8-2-15 "Side Left Cover" on page 8-51
and
8-2-16 "Side Right Cover" on page 8-52
.
2.) Remove rear side cover and front cover. Refer to the
8-2-17 "Front Cover" on page 8-53
and
3.) Unscrew 2 pcs M4 screw and 2 pcs hexa screw on the front of AC trans box. And unscrew 4 pcs
screw and 2pcs hexa screw on the rear side of Transbox assy and open inlet panel.
4.) Unplug 4 connectors inside of AC transbox and pull out the transbox assy.
5.) Pull out the transbox assy.
Figure 8-148 4 screw on the front of the AC transbox assy
2 hexa Screws