Section 4-9 - Board Function Checks
4-51
4-9-9
BackPlane Assy Function Validation Procedure
1.) Turn on the system
2.) All of attached ASSY should be tested.
3.) Refer to the
4-9-4 "SYSCONML Assy Function Validation Procedure" on page 4-47
4.) Refer to the
4-9-7 "SOM Function Validation Procedure" on page 4-50
.
5.) Refer to the
4-9-1 "CL1TRX Assy / BL1TRX Assy Function Validation Procedure" on page 4-41
6.) Refer to the
4-9-6 "P3RLY,P2RLY and P2RLY with Dummy Assy function check procedure" on page
7.) Refer to
4-9-5 "APS/APS Pro Assy PCB Function Validation Procedure" on page 4-48
.
8.) Refer to
4-9-2 "ACWD Assy Function Validation Procedure" on page 4-42
.(Optional)