4-48
Section 4-9 - Board Function Checks
4-9-5
APS/APS Pro Assy PCB Function Validation Procedure
NOTE:
HVL+ and HVL- are not available for APS PRO.
1.) Turn on the system.
2.) Do the Service Login. Refer to
5-10-2-2 "Service Login" on page 5-41
.
3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU test' => 'APS' => 'APS Overall'.
4.) Execute the diagnostic test.
5.) All diagnostic test item of APS/APS Pro should be passed.
6.) Shut down the system.
7.) Restart the system
Figure 4-47 APS/APS Pro Diagnostic Test