Section 5-10 - Service Platform
5-41
5-10-2
Global Service User Interface (GSUI)
5-10-2-1
Internationalization
The user interface provided by the service platform is designed for GE personnel and as such is in
English only. There is no multi-lingual capability built into the Service Interface.
5-10-2-2
Service Login
1.) Touch
Utility > Service
. It will take about 1p seconds for activations.
2.) Make sure that
CAPS
is OFF for password entry performed later.
3.) The Service Login window for Service Platform will be shown on the monitor display.
5-10-2-3
Access / Security
The service interface has different access and security user levels. Each user is only granted access to
the tools that are authorized for their use.
NOTE:
A Service Dongle is necessary for use by GE Service when performing proprietary level diagnostics.
OnLine Center access to the scanner requires the password and they must have ‘
Disruptive
’
permission and customer input to run diagnostics.
NOTE:
For a GE Field Engineer, the password changes at specific intervals. Access with the password is tied
to the service key.
Every access request, whether successful or not, will be logged into a service access log that is
viewable to authorized users.
Figure 5-39 Login Screen for Global Service User Interface
Table 5-5
Service Login User Levels
User Level
Access Authorization
Password
Operator
Authorized access to specified diagnostics, error logs and utilities. Same
acquisition diagnostic tests as GE Service.
uls
Administrator
uls
External Service
gogems
GE Service
Knowledge of a service level password.
A physical Service Key (Dongle) required
rotating security
password