GE H
EALTHCARE
D
IRECTION
5394141, R
EVISION
5
LOGIQ™ P5 S
EVICE
M
ANUAL
8-10
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-4
LCD Std Arm with Cover
8-2-4-1
Tools
•
Common pilIips screwdrivers
•
Allen/Unbraco wrench
8-2-4-2
Preparations
•
Shut down the system and switch off the main breaker.
•
Maneuver control console to a suitable position for removing the monitor.
8-2-4-3
Removal procedure
1.) Remove the Power Cable, DVI Cable.
2.) Unscrew 1 screw to remove the LCD arm cover.
Figure 8-13 Removing the Cable
Figure 8-14 Unscrews 1 screw
POWER CABLE
DVI CABLE