Chapter 10 Quality Assurance
10 - 7
10-5-2-2
Peripheral/Option Checks
If any peripherals or options are not part of the system configuration, the check can be omitted. Refer
to the User Manual for a list of approved peripherals/options.
10-5-3
Input Power
10-5-3-1
Mains Cable Inspection
10-5-4
Cleaning
10-5-4-1
General Cleaning
Table 10-39 GE Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1
VCR
Verify record/playback capabilities of the VCR. Clean heads and covers if necessary.
2
VGP Printer Verify hardcopy output of the video graphic printer. Clean heads and covers if necessary.
3
B/W Printer Verify hardcopy output of the B/W printer. Clean heads and covers if necessary.
4
Footswitch
Verify that the footswitch is functioning as programed. Clean as necessary.
Table 10-40 Mains Cable Inspection
Step
Item
Description
1
Unplug Cord Disconnect the mains cable from the wall and system.
2
Inspect
Inspect it and its connectors for damage of any kinds.
3
Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4
Verify
Inlet connector retainer is functional.
Table 10-41 General Cleaning
Step
Item
Description
1
Console
Use a fluid detergent in warm water on a soft, damp cloth to carefully wipe the entire system. Be careful
not to get the cloth too wet so that moisture does not enter the console.
2
Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).