GE M
EDICAL
S
YSTEMS
D
IRECTION
2317229, R
EVISION
3
LOGIQ™ 180 B
ASIC
S
ERVICE
M
ANUAL
10 - 2
Section 10-2 - Why do Periodic Maintenance
Section 10-2
Why do Periodic Maintenance
10-2-1
Keeping Records
It is good business practice that ultrasound facilities maintain records of periodic and corrective
maintenance. The Ultrasound Periodic Maintenance Inspection Certificate provides the customer with
documentation that the ultrasound scanner is maintained on a periodic basis.
A copy of the Ultrasound Periodic Maintenance Inspection Certificate should be kept in the same room
or near the scanner.
10-2-2
Quality Assurance
In order to gain accreditation from organizations such as the American College of Radiology (USA), it
is the customer’s responsibility to have a quality assurance program in place for each scanner. The
program must be directed by a medical physicists, the supervising radiologist/physician or appropriate
designee.
Routine quality control testing must occur regularly. The same tests are performed during each period
so that changes can be monitored over time and effective corrective action can be taken.
Testing results, corrective action and the effects of corrective action must be documented and
maintained on the site.
Your GE service representative can help you with establishing, performing and maintaining records for
a quality assurance program. Please contact us for coverage information and/or price for service.
Section 10-3
Periodic Maintenance Schedule
10-3-1
How often should PMs be performed?
The Care & Maintenance Task Schedule specifies how often your LOGIQ™ 180 should be serviced and
outlines items requiring special attention.
NOTE:
It is the customer’s responsibility to ensure the LOGIQ™ 180 care & maintenance is performed
as scheduled in order to retain its high level of safety, dependability and performance.
Your GE Service Representative has an in-depth knowlegde of your LOGIQ™ 180 ultrasound scanning
system and can best provide competent, efficient service. Please contact us for coverage information
and/or price for service.
The service procedures and recommended intervals shown in the Care & Maintenance Task Schedule
assumes that you use your LOGIQ™ 180 for an average patient load (10-12 per day) and use it as a
primary mobile unit which is transported between diagnostic facilities.
NOTE:
If conditions exist which exceed typical usage and patient load, then it is recommended to
increase the maintenance frequencies.