Chapter 3. Diagnostics
22
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
3.2 Displaying Diagnostic Parameters (cont.)
For example, pressing
[F1]
displays the
Tdown
parameter, as shown in the following prompt.
Table 3 below lists the available diagnostic parameters for the Model DF868. The first column in the table shows the
parameter as it appears on the option bar, while the second column shows the parameter as it appears in the prompt area
after it has been selected.
(Ch1) label SITE
Press the [
]
or [
]
keys and the appropriate
[Fx]
key to display a different
diagnostic parameter or to
EXIT
the
Diagnostics Menu
.
Ch1 DN Transit usec
519.7
(error codes appear here)
Tdown
DELTA PEAK%
Qup
Table 3: Available Diagnostic Parameters
Option Bar
Screen Display
Description
Good
Bad
SS up
UP Sig Strength
Displays signal strength for the upstream transducer.
50–75
<50 or >75
SS do
DN Sig Strength
Displays signal strength for the downstream transducer.
50–75
<50 or >75
SNDSP
Soundspeed Ft/s
Displays the measured sound speed of the fluid.
N.A.
N.A.
Tup
UP Transit S usec
Displays the upstream transit time of the ultrasonic
signal in microseconds.
N.A.
N.A.
Tdown
DN Transit usec
Displays downstream transit time of the ultrasonic
signal in microseconds.
N.A.
N.A.
DELTA
DeltaT usec
Displays the difference in
μ
sec between upstream and
downstream transit times.
N.A.
>10,000
nsec
REYN#
Reynolds #
A number based on the kinematic viscosity and flow
rate of the fluid.
N.A.
N.A.
K(RE)
K(RE)
K factor, based on Reynolds number
N.A.
N.A.
PEAK%
PEAK%
Displays the percentage of peak (set to +50 by default).
N.A.
N.A.
THETA
Theta 3 degrees
Theta 3 is the angle between the ultrasonic beam and
the normal to pipe wall. It is calculated from the
measured transit time and clamp-on parameters
(clamp-ons only).
N.A.
N.A.
Qup
UP Signal Q
Displays the signal quality for the upstream transducer.
Š 1200
–400 to +400
Qdown
DN Signal Q
Displays signal quality for the downstream transducer.
Š 1200
–400 to +400
AMPup
UP Amp Discrim
Displays the value for the amplitude discriminator of
upstream transducer.
24 ± 5
<19 or >29
AMPdn
DN Amp Discrim
Displays the value for the amplitude discriminator of the
downstream transducer.
24 ± 5
<19 or >29
CNTup
Counts up
Displays automatic gain control setting for the
upstream signal.
N.A.
N.A.