Chapter 1. Calibration
4
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
1.3.3 Calibrating the High End of the Output Range
1.
Press
[F2]
to calibrate the high end of the output range.
2.
Press
[F1]
or
[F2]
to adjust the ammeter reading
UP
or
DOWN
, until a 20 mA reading is achieved. If you press
[F3]
,
Numer
, enter the numeric value and press
[ENT]
.
3.
Press
[F4]
to
STORE
the setting or press [
and
[F1]
to
ABORT
the calibration.
Note:
If the ammeter reading cannot be adjusted within 5.0
A of the 20 mA setting, contact the factory for
assistance.
1.3.4 Testing the Analog Output Linearity
1.
Press
[F3
] to
TEST
the linearity of the currently selected analog output.
2.
Check the ammeter reading at the 50% output level.
3.
Then, enter a different output level (0-100%) and press
[ENT]
.
4.
Check the ammeter reading at this setting. Press
[ENT]
when done.
Table 1 below lists the expected ammeter readings at various
% Full Scale
settings, for both 4-20 mA and 0-20 mA
scales. Refer to this table to verify the accuracy of the ammeter readings taken above.
If the linearity test readings are not within 5
A of the values listed in Table 1 above, check the accuracy and wiring of
the ammeter. Then, repeat the low and high end calibrations. If the analog output still does not pass the linearity test,
contact the factory for assistance.
Procedure Options
You have completed calibration of the analog output. Do one of the following:
•
To calibrate or test another output, press
[F4
] (
EXIT)
and return to
Accessing the Calibration Menu,
Step 3,
•
To calibrate/test additional slots, press
[F4
] (
EXIT)
twice and proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4
]
(EXIT)
three times.
Table 1: Expected Ammeter Readings
% Full Scale
4-20 mA Scale*
0-20 mA Scale*
0%
4.000
0.000
10%
5.600
2.000
20%
7.200
4.000
30%
8.800
6.000
40%
10.400
8.000
50%
12.000
10.000
60%
13.600
12.000
70%
15.200
14.000
80%
16.800
16.000
90%
18.400
18.000
100%
20.000
20.000
* All ammeter readings should be ±0.005 mA