DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
47
Appendix B. Optional Enclosures
B.7.2 Circuit Board Installation
Refer to Figure 11 on page 50 and complete the following steps:
1.
Position the new printed circuit board within the electronics console so that it rests on the six standoffs in the base
of the enclosure and the option card sockets are located near the rear of the enclosure. Secure the circuit board in
place with the six screws previously removed.
2.
Connect the flat
LCD DISPLAY CABLE
connector to terminal J52 near the option card sockets. Make sure the marked
pin #1 side of the cable connector is located nearer the front of the enclosure.
3.
Connect the twisted-pair
DISPLAY BACKLIGHT
cable to connector
J2
beneath the left side of the printed circuit board.
This connector is polarized and can only be installed with the red wire closer to the rear of the enclosure.
4.
Connect the flat
KEYPAD CABLE
connector to terminal J50 near the front of the printed circuit board. Make sure the
marked pin #1 side of the cable connector is located nearer the rear of the enclosure.
5.
Install the
RS232
,
ANALOG OUT
and
TRANSDUCER
connectors onto the printed circuit board at the locations indicated
in Figure 11 on page 50. Make sure the marked pin #1 sides of the connectors face the right side (keypad side) of
the enclosure.
Note:
If any leads have come loose from the screw terminals on these connectors, refer to Appendix C,
Optional
Enclosures
, of the
Startup Guide
for wiring instructions.
6.
Reconnect the main
POWER LEADS
to the screw terminals at the rear of the printed circuit board. When properly
installed, the lead colors should be green, white and black, from left to right (as viewed from the front of the
enclosure).
7.
Reinstall all option cards, as described earlier in this chapter. Insert all external option card connectors at the rear
panel of the meter.
8.
After checking for any loose hardware that may have fallen into the enclosure, reinstall the top panel on the meter
and secure the panel in place with the four screws previously removed.
The Model DF868 flowmeter may now be placed back into service. Reconnect the main power and refer to Chapter 2,
Initial Setup
, of the
Startup Guide
and Chapter 1,
Calibration
, of this
Service Manual
for instructions on properly
setting up the meter for accurate flow rate measurements.
Note:
Be sure to record the printed circuit board replacement in Appendix A,
Service Record
.