Chapter 2. Error Codes and Screen Messages
18
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
E11: Poor Signal
Problem:
This message indicates a poor signal.
Cause:
The shape of the processed signal is not good enough to formulate a flow measurement.
Action:
Check the pipe and setup parameters and the transducer connections. The process may not include enough
two-phase content — switch to transit-time mode.
E12: Incoherent Signal
Problem:
This message indicates an incoherent signal.
Cause:
The flow measurements, as determined by the two Repetition Periods, are too dissimilar.
Action:
Check the pipe and setup parameters and the transducer connections. The process may not include enough
two-phase content — switch to transit-time mode.
E13: Over Range
Problem:
This error code message indicates that the present measurement exceeds the capacity of the meter.
Cause:
A internal mathematical overflow has occurred in either the volumetric or mass flow calculations.
Action:
Select larger measurement units or a shorter time interval for the current measurement parameter. For
example, choose
KSCF/M
instead of
SCF/M
. See the
Startup Guide
for instructions.
E14: Totals Overflow
Problem:
The totalizers are unable to keep up with the total flow signals.
Cause:
The programmed units/pulse value is too small.
Action:
Select a larger number of units/pulse value.
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page proceed to next page ...
Страница 22: ...Chapter 1 Calibration DigitalFlow Multipurpose Ultrasonic Liquid Flowmeter Service Manual 14 ...
Страница 64: ......
Страница 66: ......
Страница 67: ......