DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
3
Chapter 1. Calibration
1.3.1 Accessing the Calibration Menu
Figure 2: Ammeter Connection (Output A)
1.
Press the
[CAL]
key to enter the
Calibration Program.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include a slot listing for each installed option card.)
3.
Press
[Fx]
to select the desired output.
IMPORTANT:
The procedure for calibrating all outputs is the same. However, when calibrating a different output, be
sure to reconnect the ammeter to the appropriate pins on terminal block
I/O
. See Figure 1 on page 2 or
Figure 2 above for the correct pin numbers.
1.3.2 Calibrating the Low End of the Output Range
1.
Press
[F1]
to calibrate the low end of the output range.
2.
Press
[F1]
or
[F2]
to adjust the ammeter reading
UP
or
DOWN
, until a 4 mA reading is achieved. If you press
[F3]
,
Numer
, enter the numeric value and press
[ENT]
.
3.
Press
[F4]
to
STORE
the setting or press [
and
[F1]
to
ABORT
the calibration.
Note:
If the ammeter reading cannot be adjusted within 5.0
A of the 4 mA setting, contact the factory for assistance.
Terminal Block (Option Card)
B
RTN
B
SIG
A
RTN
A
SIG
8
1
+
–
Ammeter
D
RTN
D
SIG
C
RTN
C
SIG
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page proceed to next page ...
Страница 22: ...Chapter 1 Calibration DigitalFlow Multipurpose Ultrasonic Liquid Flowmeter Service Manual 14 ...
Страница 64: ......
Страница 66: ......
Страница 67: ......