Chapter 1. Calibration
12
DigitalFlow™ Multipurpose Ultrasonic Liquid Flowmeter Service Manual
1.7.3 Entering Pulse Number
5.
Enter the number of pulses desired (between 1 and 10,000) and press the
[ENT]
key. That number of pulses will
then be output at the specified frequency.
6.
Repeat steps 3, 4 and 5 to test all four of the frequency/totalizer outputs. If any of the outputs fails to pass the test,
contact the factory for assistance.
Procedure Options
You have completed testing the totalizer/frequency outputs. Do one of the following:
•
To calibrate/test additional slots, press
[F4] (EXIT)
and proceed to the appropriate section.
•
To leave the
Calibration Menu,
press
[EXIT]
twice
.
Содержание DigitalFlow DF868
Страница 2: ......
Страница 4: ...ii no content intended for this page proceed to next page ...
Страница 22: ...Chapter 1 Calibration DigitalFlow Multipurpose Ultrasonic Liquid Flowmeter Service Manual 14 ...
Страница 64: ......
Страница 66: ......
Страница 67: ......