5-6
170 Series Monitor
Revision D
2003023-001
Fetal Heart Rate Monitoring:
Fetal Heart Rate Offset
Activating the Fetal Heart Rate Offset Mode
When you activate the heart rate offset mode, the
secondary
FHR trend is shifted
+20 BPM. (See
.)
1. Ensure the recorder is
on
.
2. Press and hold the
Mark/Offse
t button
for
two seconds
. You will hear a
“beep” for confirmation.
On a Model 172 or 174 (dual ultrasound), the secondary ultrasound trace is
s20 BPM and the
symbol is printed on the upper portion
of the heart rate grid every page.
On a Model 173 or 174 (US/FECG), the ultrasound trace is s20
BPM and the
symbol is printed on the heart rate grid every page.
Upon activation, a vertical dashed line is printed to draw attention to the
start of the shifted trend.
De-Activating the Fetal Heart Rate Offset Mode
After the FHR patterns have been assessed, set the secondary FHR trend back to the
normal (unshifted) position.
1. Ensure the recorder is
on
.
NOTE:
If the auto-revert (10-MIN)
feature is enabled, the shifted heart rate
trace automatically reverts to normal
after 10 minutes. (Refer to the
“170
Series Service Manual”
for more
information.)
2. Press and hold the
Mark/Offset
button
for
two seconds
. You will hear a
“beep” for confirmation.
The trend returns to the unshifted position.
A vertical dashed line is printed to draw attention to the change.
US2+20
US + 20
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 12: ......
Страница 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Страница 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Страница 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Страница 111: ......