6-4
170 Series Monitor
Revision D
2003023-001
Uterine Activity Monitoring:
Tocotransducer (External Method)
More About Referencing
Out of Range Condition
After you press the
UA Reference
button, if there is insufficient range to provide at
least 100 relative units above the reference level (probably because the belt is too
tight), the UA display area flashes “– –”. If this happens, remove the tocotransducer
from the patient; re-reference with no pressure applied to the button; re-apply the
transducer to approximately 25 relative units
above
the baseline; then re-reference
one more time. If you still receive the same result, try a different tocotransducer or
contact your Service Representative.
Manually Setting the Baseline at the Default Value
Briefly pressing the
UA Reference
button sets the baseline at the
default
setting—the
default is configured via the service setup mode. The monitor is shipped from the
factory with a
default
setting of 10 relative units. Qualified service personnel can
access the service setup mode to set the baseline
default
to 5, 10, 15, 20, or 25
relative units. Refer to the
“170 Series Service Manual”
for more information.
Manually Overriding the Baseline Default Value
Pressing and holding the
UA Reference
button for more than two seconds causes the
UA reference level and display to override the default setting and cycle through all
available selections: 5, 10, 15, 20, or 25 relative units, starting at the default
setting—until the button is released. Once the button is released, the UA trace and
UA value take on this new value as a baseline for reference.
Briefly pressing the
UA Reference
button
reverts back to using the
default
setting configured via the service setup mode.
Automatic Baseline “Zeroing”
If pressure falls below 0 relative units (probably because the belt has loosened),
automatic UA referencing occurs and a new baseline reference is set at 0 relative
units.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 12: ......
Страница 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Страница 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Страница 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Страница 111: ......