11-4
170 Series Monitor
Revision D
2003023-001
Technical Specifications:
Strip Chart Recorder
Strip Chart Recorder
NOTE:
Specifications subject to change without notice.
Table 11-3. Strip Chart Recorder Specifications
Heart Rate Scale
Chart Width:
Scaling:
Range:
Resolution:
Domestic
International
7 cm
8 cm
30 BPM/cm
20 BPM/cm
30–240 BPM
50–210 BPM
1 BPM
1 BPM
Uterine Activity Scale
Chart Width:
Scaling:
Range:
Resolution:
St
rain Gauge
Tocotransducer
4 cm
4 cm
25 mmHg/cm
25 relative units/cm
0–100 mmHg
0–100 relative units
1 mmHg
1 relative unit
Recorder Drive
Speeds:
Speed Accuracy:
1, 2, and 3 cm/min
±2 % over 10 minutes