Revision D
170 Series Monitor
4-15
2003023-001
Setup Procedures:
Preparing the Monitor for Patient Use
Preparing the Monitor for Patient Use
The following steps should be performed prior to each patient monitoring session:
1. Ensure an adequate supply of paper is in the recorder. The recorder
automatically stops when paper runs out. If the recorder requires paper, refer to
.
2. Press the
Power
button and wait for the monitor to complete the self-test routine.
Refer to
. Ensure the monitor enters
the normal operating mode.
NOTE:
if you need to
change an alarm or time/date setting.
3. Check the status of the monitor’s alarms disabled indicator. When lit the alarms
are disabled.
4. If the alarms are
enabled
, check the strip chart printout and acknowledge the
settings for the high and low alarm limits for the fetal heart rate.
5. Check the time and date that is printed on the strip chart paper.
6. Attach the appropriate transducer(s) for monitoring. Read the
“Maternal/Fetal
Monitoring Operator’s Manual”
for instructions on applying the transducers.
7. Press the
Record
button and ensure that the chart paper moves freely out the
front of the recorder drawer.
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 12: ......
Страница 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Страница 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Страница 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Страница 111: ......