5-10
170 Series Monitor
Revision D
2003023-001
Fetal Heart Rate Monitoring:
Fetal Heart Rate Alarms
Latching Alarms
Fetal heart rate threshold alarms are “latching.” This means that a clinician must
acknowledge the alarm using the monitor’s
Alarm Silence
button in order to clear the
alarm.
Active Threshold Alarm:
Press the
Alarm Silence
button
to cancel the audio
component of an active threshold alarm. The visual indications remains present
until the FHR value returns to within the defined acceptable range.
Unsilenced, Resolved Threshold Alarm:
If a threshold alarm condition
resolves, prior to being silenced (clinical acknowledgment), the visual and
audible indications both remain present. Press the
Alarm Silence
button
to
cancel both the audible and visual indications.
FHR High Alarm
The simplest example of a high FHR alarm occurs when the FHR value is
continuously
greater than the threshold (high limit) for 5 minutes. When data
consistently violates the limit, the time-to-alarm is 5 minutes. See
.
Figure 5-5. High FHR Alarm Example
1
2
3
4
5
Time (minutes)
FHR (BPM)
high alarm limit
(threshold)
alarm activates
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series OPERATOR S MANUAL MANUAL P N 2003023 001 REV D ...
Страница 12: ......
Страница 21: ...Revision D 170 Series Monitor 1 9 2003023 001 Safety Definitions of Symbols For your notes ...
Страница 22: ...1 10 170 Series Monitor Revision D 2003023 001 Safety Definitions of Symbols ...
Страница 42: ...3 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 58: ...4 16 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 78: ...6 6 170 Series Monitor Revision D 2003023 001 For your notes ...
Страница 106: ...10 4 170 Series Monitor Revision D 2003023 001 Supplies Accessories ...
Страница 111: ......